The ISA-60M controller with remote MRI-5175 oxygen sensor/transmitter is specifically designed to continuously monitor for oxygen deficiency in a Magnetic Resonance Imaging (MRI) facility. Maintaining the quality of breathing air in an MRI room is an important safety consideration. Cryogenic gas used to maintain the magnetic field is highly dangerous if released into an enclosed space. Inert gas can create a hazardous situation by reducing the concentration of oxygen in the breathing air to unsafe levels. The ISA-60M system ensures the safety of the technicians and patients by continuously monitoring and displaying the levels of oxygen in the room and providing audio and visual alarms when there is an unsafe, low level oxygen condition in the area being monitored. Three verisons available; Single Channel, Dual Channel, and Three Channel.

Product Highlights

  • Continuous monitoring of area with Long-life oxygen sensor
  • Oxygen sensor/transmitter is designed for EMC compatibility
  • User-programmable alarms and high capacity relays
  • 4-20 mA inputs and outputs
  • 100-240 VAC, 12 VDC power options
  • Fault relay and visual fault alarm
  • RS-232/485 with Modbus protocol
  • Safety Approvals: CE
  • Conforms to STD 61010-1
  • Certified to C22.2 #61010-1

 

Unit of Measure

Financing

Approve Payments

N/A Finance with Approve Payments

Specifications

Alarm Level Setting

N/A Alarm 1: 5 ppm CO, user adjustable Alarm 2: 10 ppm CO, user adjustable Alarm 3: 20 ppm CO, user adjustable, All adjustable between 1 to 50 ppm

Audible Alarm

N/A 105 dB at 12 in 95 dB at 24 in

Inlet Size

N/A Strain relief (2 supplied) for 0.20” to 0.35” OD Cable

Operating Temperature Range

N/A 32° to 104°F ºF

Power Requirements

N/A 100 to 240 VAC and/or 12 VDC, 15W 12 VDC

Weight

N/A 10 lbs

General Description

N/A

Temperature: 32° to 104°F (0° to 40°C)

 

Relative Humidity: 10-99% RH, non-condensing

 

Pressure, Atmospheric: 20 to 36 inHg (68 to 122 kPa)

 

Approval: Safety Approvals: CE, Conforms to STD 61010-1, Certified to C22.2 #61010-1

 

 

 

ISA-60M MONITOR SPECIFICATIONS:

 

Standard Alarm Points: 19.5% & 17% decreasing oxygen

 

Display: High-contrast, backlit

 

Alarms: Visual: LEDs; Audible: Piezoelectric horn, 95 dB at 2 feet

 

Alarm Relays: 5 programmable gas relays plus fault. All relays are programmable latching or non-latching, dry SPDT, 10 amps (resistive load only) at 110 VAC.

 

Operating Power: 100 to 240 VAC and/or 12 VDC, 15 Watts

 

Connections: Strain relief (2 supplied) for 0.20” to 0.35” OD Cable

 

Operating Power: 100 to 240 VAC and/or 12 VDC, 15 Watts

 

Enclosure: Thermoplastic box with clear, hinged front cover, designed for NEMA 12 and 4X

 

Size: 10.5 H x 8.5 W x 5.8 D inches (27 x 22 x 15 cm)

 

Weight: 8 lbs. (3.6 kg)

 

 

 

MRI-5175 TRANSMITTER:

 

Alarms: Visual: 3 Red LEDs; Visual Power/Fault: Green/Red LED; Optional Audible Alarm: 105 dB at 12 inches (30 cm)

 

Outputs: 4-20 mA, RS-485 Modbus

 

Power: 24 VDC from ISA-60M

 

Connection to ISA-60M: Strain relief (1 supplied) for 24 VDC, 4-20 mA, 0.20” to 0.35” OD Cable

 

Enclosure: Fiberglass-reinforced polyester with non-magnetic coating and hardware

 

Dimensions: 5 x 5 x 4 inches (2 x 2 x 1.6 cm)

 

Weight: 2 lbs. (0.9 kg)

 

Oxygen Sensor supplied with MRI-5175: Type: Electrochemical; Range: 0 to 30.0% by Vol. O2; Resolution: 0.1% by Volume Oxygen; Typical Life: 4 to 5 years

 

Sensor Cable from MRI-5175 to ISA-60M: 3-wire, shielded cable, typically 18 to 20 AWG, user supplied, UL Listed type CM, PLTC or TC