The ISA-60M controller with remote MRI-5175 oxygen sensor/transmitter is specifically designed to continuously monitor for oxygen deficiency in a Magnetic Resonance Imaging (MRI) facility. Maintaining the quality of breathing air in an MRI room is an important safety consideration. Cryogenic gas used to maintain the magnetic field is highly dangerous if released into an enclosed space. Inert gas can create a hazardous situation by reducing the concentration of oxygen in the breathing air to unsafe levels. The ISA-60M system ensures the safety of the technicians and patients by continuously monitoring and displaying the levels of oxygen in the room and providing audio and visual alarms when there is an unsafe, low level oxygen condition in the area being monitored. Three verisons available; Single Channel, Dual Channel, and Three Channel.
Product Highlights
Alarm Level Setting |
N/A Alarm 1: 5 ppm CO, user adjustable Alarm 2: 10 ppm CO, user adjustable Alarm 3: 20 ppm CO, user adjustable, All adjustable between 1 to 50 ppm |
Audible Alarm |
N/A 105 dB at 12 in 95 dB at 24 in |
Inlet Size |
N/A Strain relief (2 supplied) for 0.20” to 0.35” OD Cable |
Operating Temperature Range |
N/A 32° to 104°F ºF |
Power Requirements |
N/A 100 to 240 VAC and/or 12 VDC, 15W 12 VDC |
Weight |
N/A 10 lbs |
General Description |
N/A
Temperature: 32° to 104°F (0° to 40°C)
Relative Humidity: 10-99% RH, non-condensing
Pressure, Atmospheric: 20 to 36 inHg (68 to 122 kPa)
Approval: Safety Approvals: CE, Conforms to STD 61010-1, Certified to C22.2 #61010-1
ISA-60M MONITOR SPECIFICATIONS:
Standard Alarm Points: 19.5% & 17% decreasing oxygen
Display: High-contrast, backlit
Alarms: Visual: LEDs; Audible: Piezoelectric horn, 95 dB at 2 feet
Alarm Relays: 5 programmable gas relays plus fault. All relays are programmable latching or non-latching, dry SPDT, 10 amps (resistive load only) at 110 VAC.
Operating Power: 100 to 240 VAC and/or 12 VDC, 15 Watts
Connections: Strain relief (2 supplied) for 0.20” to 0.35” OD Cable
Operating Power: 100 to 240 VAC and/or 12 VDC, 15 Watts
Enclosure: Thermoplastic box with clear, hinged front cover, designed for NEMA 12 and 4X
Size: 10.5 H x 8.5 W x 5.8 D inches (27 x 22 x 15 cm)
Weight: 8 lbs. (3.6 kg)
MRI-5175 TRANSMITTER:
Alarms: Visual: 3 Red LEDs; Visual Power/Fault: Green/Red LED; Optional Audible Alarm: 105 dB at 12 inches (30 cm)
Outputs: 4-20 mA, RS-485 Modbus
Power: 24 VDC from ISA-60M
Connection to ISA-60M: Strain relief (1 supplied) for 24 VDC, 4-20 mA, 0.20” to 0.35” OD Cable
Enclosure: Fiberglass-reinforced polyester with non-magnetic coating and hardware
Dimensions: 5 x 5 x 4 inches (2 x 2 x 1.6 cm)
Weight: 2 lbs. (0.9 kg)
Oxygen Sensor supplied with MRI-5175: Type: Electrochemical; Range: 0 to 30.0% by Vol. O2; Resolution: 0.1% by Volume Oxygen; Typical Life: 4 to 5 years
Sensor Cable from MRI-5175 to ISA-60M: 3-wire, shielded cable, typically 18 to 20 AWG, user supplied, UL Listed type CM, PLTC or TC |