MedAir 2200 is a compact, easy to install and operate compressed airline monitor that can continuously and simultaneously monitor up to four gases in real-time using internally mounted sensors. MedAir 2200 is typically configured to monitor hospital compressed air for carbon monoxide, oxygen and dew point. Four versions available CO Monitor, CO and Oxygen Monitor, CO and Dew Point, and  CO, Oxygen and Dew Point.

Product Highlights

  • Continuous real-time gas monitoring of hospital compressed airlines
  • Monitors oxygen, carbon monoxide and dew point simultaneously
  • User programmable alarms and relays
  • Meets Grade D breathing and NFPA 99 requirements
  • Low flow alarm
  • Field replaceable sensors

Link







Financing  · Specifications

Financing

Approve Payments

Finance with Approve Payments



Specifications

Range (CO)

0 - 50 ppm

Accuracy

0-30 %.

Alarm Level Setting

10 and 20 ppm CO
Alarm 1: +35°F Dew Point

Alarm 2: +50°F Dew Point

Audible Alarm

95 dB at 24 in

Inlet Size

1/4" NPT Female

Power Requirements

100 to 240 VAC and/or 12 VDC, 15W

Weight

8 lbs

General Description

Display: High-contrast, backlit

 

Alarms: Visual: LEDs

 

Audible: Piezoelectric horn, 95 dB at 2 feet

 

Alarm Relays: 5 programmable gas relays plus fault. All relays are programmable latching or nonlatching, dry SPDT, 10 amps (resistive load only) at 110 VAC

 

Operating Power: 100 to 240 VAC and/or 12 VDC, 15 Watts

 

Sensor Types:   CO and O2: Electrochemical, Dew Point: Thin-Film Polymer

 

Sensor Life: CO: 1-3 years, with calibration, O2: 1-2 years, with calibration, Dew Point: 2 years, replacement

 

Response Times: Dew point: T90 = 10 seconds for -40° to 50°F step change, CO: T90 = 30 seconds, O2: T90 = 15 seconds

 

Inlet Pressure: Must be regulated to 55 psi (user provided)

 

Flow Rate: 1/2 liter per minute (1 SCFH)

 

Enclosure: Thermoplastic box with clear, hinged front cover, designed for NEMA 12 and 4X

 

Dimensions: 9.3 W x 11.9 H x 6.1 D inches (21.3 W x 30.3 H x 15.4 D cm)

 

Weight: 8 lbs. (3.6 kg)

 

Approvals: UL 60601-1, IEC 60601-1 & CSA 22.2, No. 601.1

 

 

 

STANDARD CALIBRATIONS

 

Range: 0-50 ppm CO, 0-30% O2 Deficiency, 0-30% O2 Abundance, -112° to +68°F Dew Point(-80° to +20°C)

 

Alarm Points: 10 and 20 ppm CO, 19.5% O2 Deficiency, 23.5% O2 Abundance, Alarm 1: +35°F Dew Point (+1.6° C), Alarm 2: +50°F Dew Point (+10° C)

X

Display: High-contrast, backlit

 

Alarms: Visual: LEDs

 

Audible: Piezoelectric horn, 95 dB at 2 feet

 

Alarm Relays: 5 programmable gas relays plus fault.  All relays are programmable latching or nonlatching, dry SPDT, 10 amps (resistive load only) at 110 VAC

 

Operating Power: 100 to 240 VAC and/or 12 VDC, 15 Watts

 

Sensor Types:    CO and O2: Electrochemical, Dew Point: Thin-Film Polymer

 

Sensor Life: CO: 1-3 years, with calibration, O2: 1-2 years, with calibration, Dew Point: 2 years, replacement

 

Response Times: Dew point: T90 = 10 seconds for -40° to 50°F step change, CO: T90 = 30 seconds, O2: T90 = 15 seconds

 

Inlet Pressure: Must be regulated to 55 psi (user provided)

 

Flow Rate: 1/2 liter per minute (1 SCFH)

 

Enclosure: Thermoplastic box with clear, hinged front cover, designed for NEMA 12 and 4X

 

Dimensions: 9.3 W x 11.9 H x 6.1 D inches (21.3 W x 30.3 H x 15.4 D cm)

 

Weight: 8 lbs. (3.6 kg)

 

Approvals: UL 60601-1, IEC 60601-1 & CSA 22.2, No. 601.1

 

 

 

STANDARD CALIBRATIONS

 

Range: 0-50 ppm CO, 0-30% O2 Deficiency, 0-30% O2 Abundance, -112° to +68°F Dew Point(-80° to +20°C)

 

Alarm Points: 10 and 20 ppm CO, 19.5% O2 Deficiency, 23.5% O2 Abundance, Alarm 1: +35°F Dew Point (+1.6° C), Alarm 2: +50°F Dew Point (+10° C)