Dimension A |
N/A 39 in |
Dimension B |
N/A 68 in |
Dimension C |
N/A 95 in |
Inlet Size |
N/A 2 in |
Outlet Size |
N/A 2 in |
Air Flow at 19 inHg |
N/A 74 scfm |
Pump Power |
N/A 5 hp |
Tank Capacity |
N/A 60 gal |
Tank Orientation |
N/A Horizontal |
Capacity |
N/A 21676 Btu/hr |
Noise Level |
N/A 77 dBA |
System Full Load Current (F.L.A) at 208 V |
N/A 30 A |
System Full Load Current (F.L.A) at 230 V |
N/A 29.6 A |
System Full Load Current (F.L.A) at 460 V |
N/A 15.8 A |
System Weight |
N/A 1,815 lb |
Brands |
N/A Powerex® |
Type |
N/A Duplex |
Warranty (Limited) |
N/A 3 years |
Electrical Panel Opening |
N/A 42 in |
Minimum Clearance for Ease of Maintenance |
N/A 24 in |
Motor Voltage |
N/A 208 to 230 V460 V |
Phase |
N/A 3 |
Power Supply Frequency |
N/A 60 Hz |
Innovative Technology |
N/A Laboratory Oil Sealed Rotary Vane Systems (with HMI, PBMI or Basic controls) |
Control Panel Protection |
N/A NEMA 12 NFPA 99 |
Industry Standards/Certifications |
N/A ASME ISO 9001 NFPA UL |
N/A
The Powerex® laboratory vacuum system is designed to create a suction system to remove unwanted fluids or gases from laboratory working areas. The laboratory vacuum system package is compliant with the NFPA 99 requirements for Risk Category 3 laboratory systems. |
N/A
Each vacuum pump shall be of the rotary vane air-cooled design. Each vacuum pump shall be direct driven through a shaft coupling by a TEFC electric motor. Belt drives shall not be permitted. The vacuum pumps shall be mounted on vibration isolators. |
N/A The motor is continuous duty, C-face, TEFC, suitable for 208-230, or 460 V, 3 phase, 60 hertz electrical operation. |
N/A The system shall include an ASME rated air receiver. The tank shall be internally lined and come with a manual drain and sight gauge. Systems with Basic Controls are also equipped with a vacuum gauge. Receiver is plumbed so that system vacuum must pass through the air receiver prior to the pumps to provide additional protection for the pumps. |
N/A
The control system provides automatic lead/lag sequencing and automatic alternation of all pumps in order to equalize the amount of usage among the available vacuum pumps. The Premium NFPA Control Panel shall include a gateway server card and all features listed below: |
Service |
N/A
Powerex® has an extensive network of trained sales and service dealers - over 220 authorized service centers throughout the United States, Canada, the Middle East, Central and South Americas. In order to maintain a high level of service, we offer our distributors the chance to participate in our in-house training sessions to be up-to-date on our product offerings, installation and start-up procedures and troubleshooting solutions. |
Engineering |
N/A
Due to our vertical integration and full in-house engineering staff, Powerex® has the ability to customize both electrical control panels as well as the mechanical portion of the system to meet a wide variety of customer needs. Thanks to our continuous process improvement policy and how easily our Sales, Customer Service, Engineering and Production teams can communicate with each other and our end customers, we are constantly figuring out ways to improve our current products and add new products to our standard offerings. |
Innovation |
N/A
Powerex® was the first company to package NFPA 99 compliant scroll systems. Today, we design and manufacture multiple NFPA 99 compliant product lines - Vacuum, Air, and Instrument Air. We were the first company to utilize composite piston technology, now standard on all Powerex® reciprocating compressors. Powerex® also introduced the first dry scroll to the US oil-less compressor market in 1989. The scroll pump’s completely dry compressor chamber along with the low noise and minimal vibration is perfect for applications or locations not considered with reciprocating compressors. |
Quality |
N/A
Powerex® is a highly integrated operation where we not only assemble systems, but we also manufacture many of the major components. From compressors, UL control panels, desiccant dryers, monitors to air receivers, we control quality at every stage of design and manufacturing. All of our facilities are ISO 9001 certified. Unlike our competition, we have used the same compressor brand for decades and also retain the same company ownership since our inception. This allows us to maintain quality and provide consistency in our product offerings year after year. |
N/A
|
N/A
|
N/A Powerex® offers three different control panel options for our Laboratory Vacuum Systems - HMI controls as our standard and optional PBMI or Basic controls. |
N/A
|
N/A
|
N/A
|
N/A
|
N/A
|
N/A A rotor is mounted eccentrically in the pump cylinder and contains several sliding vanes. As the rotor turns, centrifugal force causes the vanes to slide outward against the cylinder wall. Oil is injected into the pumping chamber to create a seal between the vanes and the cylinder wall and to lubricate the vanes for reduced wear. As a result of the offset rotor, a succession of variable volumes is formed in the cylinder housing creating the flow of air through the pump. Air is pulled into the pump inlet which is then compressed and discharged into the exhaust box. At this point oil is mixed with the air and is passed through several stages of internal oil and mist eliminators to remove 99.9% of the lubricating oil before the exhaust is released to atmosphere. The separated oil is then returned to the oil reservoir. |
N/A
Optional Auto Purge |
N/A
This option is available for systems with Standard HMI controls or Optional PBMI controls. It is not available on systems with Basic controls. |