Dimensions

Length (A)

N/A 55 in

Width (B)

N/A 66 in

Height (C)

N/A 76 in

Inlet Connection Size

N/A 2 in

Outlet Connection Size

N/A 1 in

Tankmount Vacuum Systems

Pump Power

N/A 4 hp
Air Flow at 19 inHg1 N/A 29 scfm

Tank Capacity

N/A 120 gal

Tank Orientation

N/A Vertcal

Compressor Capacity

N/A 8,670 Btu/hr

Sound Level

N/A 79 dBA

System Full Load Current (F.L.A) at 208 V

N/A 22 A

System Full Load Current (F.L.A) at 230 V

N/A 20.4 A

System Full Load Current (F.L.A) at 460 V

N/A 11.2 A

System Weight

N/A 2,200 lb

Specifications

Electrical Panel Opening

N/A 42 in

Minimum Clearance for Ease of Maintenance

N/A 24 in

Motor Voltage

N/A 208 V230 V460 V

Phase

N/A 3

Power Supply Frequency

N/A 60 Hz

Vacuum System

N/A The Powerex medical vacuum system is designed to create a suction system to remove unwanted fluids or gases from hospital/ laboratory working areas. The medical vacuum system package is compliant with the NFPA 99 requirements for Risk Category 1 systems. Each system is completely tested before shipment and includes:

• Multiple vacuum pumps and associated equipment.
• AMSE air receiver.
• Medical control panel. Each pump is factory piped to a common intake manifold. Vibration isolation pads are included with the system.

Rotary Vane Vacuum Pump

N/A Each pump shall be a rotary claw type vacuum pump, and shall be direct-driven through a shaft coupling by a C-face, TEFC electric motor.

• Each vacuum pump shall be dry-running, featuring two claw-type, non-contacting rotors and shall not require any sealing fluid in the pumping chamber, assuring virtually maintenance-free operation.
• Each vacuum pump shall include an internal relief valve, and a built-in, anti-suck-back valve mounted at the pump inlet.
• Each pump within the system shall include a check valve, inlet and discharge flex connectors, a 5 micron inlet filter and a pump isolation valve.

Motor

N/A The motor is continuous duty, C-face, TEFC, suitable for 208-230, or 460V, 3 phase, 60 hertz electrical operation.

Receiver

N/A The system shall include an ASME rated air receiver. The tank shall be equipped with a vacuum gauge, a sight gauge, by-pass valves, and a manual drain.

Control Panel

N/A The control system provides automatic lead/lag sequencing and automatic alternation of all pumps in order to equalize the amount of usage among the available vacuum pumps. The Premium NFPA Control panel shall include a gateway server card and all features listed below:

• PLC controller and a color touch screen panel which displays the operating status of the unit.
• Building automation communication gateway with BacNet® protocol and Web server features. Web servers features include email notifications in case the system is in alarm or has achieved one its maintenance intervals and requires service.
• Ethernet port for connection to BacNet® server or direct connection to facility Ethernet for viewing of system operations and status via device IP-address.
• UL508A listed control panel in a NEMA 12 enclosure. The panel door will include: the HMI touch screen, an audible and visual alarms with an acknowledge button, and an HOA switch for each pump.
• Magnetic starters.
• Vacuum transducer for process control.
• Single point power connection.
• Redundant 120Vac control transformers with fused primary and secondary protection.
• System overload trip, high temperature conditions or maintenance intervals for the pump will result in visual and audible alarms.

Available Options

N/A • Oxygen assured pumps - prepared for use in WAGD systems.
• Internal tank lining for corrosion resistance.
• Variable Frequency Drive (VFD) Control Panel – with Premium NFPA controls.

Note

N/A 3 Year Limited Warranty

  • 1 Air Flow shown at total system capacity.