LSD02A3 2HP Laboratory Open Oilless Scroll Duplex Air Compressor with Premium NFPA Controls |
LSD03A3 3HP Laboratory Open Oilless Scroll Duplex Air Compressor with Premium NFPA Controls |
LSD05A3 5HP Laboratory Open Oilless Scroll Duplex Air Compressor with Premium NFPA Controls |
LSD10A4 10HP Laboratory Open Oilless Scroll Duplex Air Compressor with Premium NFPA Controls |
LST05A4 5HP Laboratory Open Oilless Triplex Air Compressor with Premium NFPA Controls |
|||||
Description | N/A The Powerex® Laboratory Open Scroll Air Compressor System is designed to provide clean, dry air for applications where the quality of the compressed air is critical. The standard unit is rated for a maximum of 145 psig. | |||||||||
Dimension A | N/A 32 in | |||||||||
Dimension B | N/A 74 in | N/A 74 in | N/A 83 in | N/A 83 in | N/A 83 in | |||||
Dimension C | N/A 62 in | N/A 62 in | N/A 68 in | N/A 77 in | N/A 77 in | |||||
Optional Inlet (X2) | N/A 3/4 | N/A 3/4 | N/A 3/4 | N/A 1 1/2 | N/A 1 | |||||
Outlet Size | N/A 1/2 in | N/A 1/2 in | N/A 1/2 in | N/A 1 in | N/A 1 in | |||||
Power | N/A 2 hp | N/A 3 hp | N/A 5 hp | N/A 10 hp | N/A 5 hp | |||||
Type | N/A Duplex | N/A Duplex | N/A Duplex | N/A Duplex | N/A Triplex | |||||
Air Flow at 145 psig | N/A 9.2 scfm | N/A 14.2 scfm | N/A 25 scfm | N/A 50 scfm | N/A 37.5 scfm | |||||
Tank Capacity | N/A 80 gal | N/A 80 gal | N/A 80 gal | N/A 120 gal | N/A 120 gal | |||||
Air Flow | N/A 9740 Btu/hr | N/A 14600 Btu/hr | N/A 24400 Btu/hr | N/A 48800 Btu/hr | N/A 36600 Btu/hr | |||||
Sound Level | N/A 70 dBA | N/A 75 dBA | N/A 75 dBA | N/A 77 dBA | N/A 76 dBA | |||||
Warranty (Limited) | N/A 3 years | |||||||||
System Full Load Current (F.L.A) at 208 V | N/A 15 A | N/A 21.2 A | N/A 33.4 A | N/A 66.8 A | N/A 50.1 A | |||||
System Full Load Current (F.L.A) at 230 V | N/A 13.6 A | N/A 19.2 A | N/A 30.4 A | N/A 60.8 A | N/A 45.6 A | |||||
System Full Load Current (F.L.A) at 460 V | N/A 6.8 A | N/A 9.6 A | N/A 15.2 A | N/A 30.4 A | N/A 22.8 A | |||||
System Weight | N/A 1,300 lb | N/A 1450 lb | N/A 1,500 lb | N/A 1,900 lb | N/A 1,850 lb | |||||
Electrical Panel Opening | N/A 42 in | |||||||||
Minimum Clearance for Ease of Maintenance | N/A 24 in | |||||||||
Brands | N/A Powerex® | |||||||||
Open Scroll Air Compressor System |
N/A
Each compressor shall be belt driven oil-less rotary scroll single stage, air-cooled construction with absolutely no oil needed for operation. Direct drive compressors shall not be used.
|
|||||||||
Motor | N/A Each compressor shall be belt driven by a 1750 RPM, ODP, NEMA construction motor. Motors running at speeds higher than 1750 RPM shall not be acceptable. OSHA approved belt guards shall be provided. | |||||||||
Air Receiver |
N/A
The system shall include an ASME air receiver rated for 200 PSI MAWP. The tank shall be equipped with:
|
|||||||||
Premium NFPA Control Panel |
N/A
The control system provides automatic lead/lag sequencing and automatic alternation of all compressors in order to equalize the amount of usage among the available compressors. The Premium NFPA Control Panel shall include a gateway server card and all features listed below:
|
|||||||||
Air Purification Package | N/A The air purification package shall consist of the following: dual desiccant air dryers, dual filter and regulator bank with sample ports, and dew point monitor with alarms, and all bypass piping. Piping to be brass, stainless, or type K copper, and cleaned for medical air use. All components shall be mounted piped and wired to the air receiver. | |||||||||
Desiccant Air Dryers |
N/A
The twin-tower, heatless desiccant air dryer is designed for a pressure dew point of -40 degrees F. Each dryer is sized to meet the full SCFM requirement of the system. The dryer shall be installed and plumbed on a common steel skid to the air receiver. The dryer design shall be of the automatic pressure swing, heatless, regenerative type and shall include a solid state cycle timer, reliable switching valve design, and purge muffler for quiet operation. Dryer controls shall include on/off/variable mode selector switch, and digital display. The filtration system shall consist of 2 stages of filtration mounted and plumbed to the air dryer. The first stage of filtration shall include a centrifugal water separation and a 5 micron filter element with element coalescing vibration or a .01 micron or finer pre-filter with element change indicator and automatic condensate drain and installed upstream of the air dryer. The second stage shall include a 1 micron (or finer) particulate filter with element change indicator and installed downstream of the air dryer. |
|||||||||
Filtration And Pressure Reducing Station |
N/A
Each filter/dryer/regulator assembly shall be plumbed with bypass valves to enable service without disrupting air flow to the facility. Each assembly also includes a sample air port. The filtration systems consist of two stages of filtration as outlined above in the Desiccant Air Dryers section. Additionally, each assembly shall include:
|
|||||||||
Additional Information |
N/A
Each system is completely tested before shipment and includes:
|
|||||||||
Note |
N/A
For Air Flow at 19 inHg, Compressor Capacity, System Full Load Current (F.L.A) at 208 V, System Full Load Current (F.L.A) at 230 V, and System Full Load Current (F.L.A) at 460 V values listed are with all pumps running. 10 hp systems will have one separator per two pumps. |
|||||||||
|