Medical Claw Open Systems
The claw pump design is extremely efficient and has very low maintenance cost; because there is no actual contact between the rotors and the cylinder wall, parts that need replacing or required maintenance is minimized.
Benefit & Features
Items |
CVTD0203 2 HP Duplex Tankmount Medical Claw Open Vaccum System CVTD0203 |
CVTD0203V 2 HP Duplex Tankmount Medical Claw Open Vaccum System CVTD0203V |
CVTD0303 3 HP Duplex Tankmount Medical Claw Open Vaccum System CVTD0303 |
CVTD0303V 3 HP Duplex Tankmount Medical Claw Open Vaccum System CVTD0303V |
CVTD0504A 4 HP Duplex Tankmount Medical Claw Open Vaccum System CVTD0504A |
|||||
Length (A) | N/A 71 in | N/A 71 in | N/A 71 in | N/A 71 in | N/A 77 in | |||||
Width (B) | N/A 42 in | N/A 42 in | N/A 42 in | N/A 42 in | N/A 43 in | |||||
Height (C) | N/A 68 in | N/A 68 in | N/A 68 in | N/A 68 in | N/A 72 in | |||||
Inlet Connection Size | N/A 1.5 in | N/A 1.5 in | N/A 1.5 in | N/A 1.5 in | N/A 2 in | |||||
Outlet Connection Size | N/A 1 in | |||||||||
Pump Power | N/A 2 hp | N/A 2 hp | N/A 3 hp | N/A 3 hp | N/A 4 hp | |||||
Air Flow at 19 inHg1 | N/A 16 scfm | N/A 16 scfm | N/A 21 scfm | N/A 21 scfm | N/A 29 scfm | |||||
Tank Capacity | N/A 80 gal | N/A 80 gal | N/A 80 gal | N/A 80 gal | N/A 120 gal | |||||
Tank Orientation | N/A Horizontal | N/A Vertcal | N/A Horizontal | N/A Vertcal | N/A Horizontal | |||||
Compressor Capacity | N/A 4,335 Btu/hr | N/A 4,335 Btu/hr | N/A 6,503 Btu/hr | N/A 6,503 Btu/hr | N/A 8,670 Btu/hr | |||||
Sound Level | N/A 70 dBA | N/A 70 dBA | N/A 70 dBA | N/A 70 dBA | N/A 79 dBA | |||||
System Full Load Current (F.L.A) at 208 V | N/A 14.4 A | N/A 14.4 A | N/A 20.8 A | N/A 20.8 A | N/A 22 A | |||||
System Full Load Current (F.L.A) at 230 V | N/A 13.2 A | N/A 13.2 A | N/A 18.8 A | N/A 18.8 A | N/A 20.4 A | |||||
System Full Load Current (F.L.A) at 460 V | N/A 7.6 A | N/A 7.6 A | N/A 10.4 A | N/A 10.4 A | N/A 11.2 A | |||||
System Weight | N/A 1000 lb | N/A 1000 lb | N/A 1,200 lb | N/A 1,200 lb | N/A 1,400 lb | |||||
Brands | N/A | |||||||||
Electrical Panel Opening | N/A 42 in | |||||||||
Minimum Clearance for Ease of Maintenance | N/A 24 in | |||||||||
Motor Voltage | N/A 208 V230 V460 V | |||||||||
Phase | N/A 3 | |||||||||
Power Supply Frequency | N/A 60 Hz | |||||||||
Industry Standards/Certifications | N/A | |||||||||
Vacuum System |
N/A
The Powerex medical vacuum system is designed to create a suction system to remove unwanted fluids or gases from hospital/ laboratory working areas. The medical vacuum system package is compliant with the NFPA 99 requirements for Risk Category 1 systems. Each system is completely tested before shipment and includes: • Multiple vacuum pumps and associated equipment. • AMSE air receiver. • Medical control panel. Each pump is factory piped to a common intake manifold. Vibration isolation pads are included with the system. |
|||||||||
Rotary Vane Vacuum Pump |
N/A
Each pump shall be a rotary claw type vacuum pump, and shall be direct-driven through a shaft coupling by a C-face, TEFC electric motor. • Each vacuum pump shall be dry-running, featuring two claw-type, non-contacting rotors and shall not require any sealing fluid in the pumping chamber, assuring virtually maintenance-free operation. • Each vacuum pump shall include an internal relief valve, and a built-in, anti-suck-back valve mounted at the pump inlet. • Each pump within the system shall include a check valve, inlet and discharge flex connectors, a 5 micron inlet filter and a pump isolation valve. |
|||||||||
Motor | N/A The motor is continuous duty, C-face, TEFC, suitable for 208-230, or 460V, 3 phase, 60 hertz electrical operation. | |||||||||
Receiver | N/A The system shall include an ASME rated air receiver. The tank shall be equipped with a vacuum gauge, a sight gauge, by-pass valves, and a manual drain. | |||||||||
Control Panel |
N/A
The control system provides automatic lead/lag sequencing and automatic alternation of all pumps in order to equalize the amount of usage among the available vacuum pumps. The Premium NFPA Control panel shall include a gateway server card and all features listed below: • PLC controller and a color touch screen panel which displays the operating status of the unit. • Building automation communication gateway with BacNet® protocol and Web server features. Web servers features include email notifications in case the system is in alarm or has achieved one its maintenance intervals and requires service. • Ethernet port for connection to BacNet® server or direct connection to facility Ethernet for viewing of system operations and status via device IP-address. • UL508A listed control panel in a NEMA 12 enclosure. The panel door will include: the HMI touch screen, an audible and visual alarms with an acknowledge button, and an HOA switch for each pump. • Magnetic starters. • Vacuum transducer for process control. • Single point power connection. • Redundant 120Vac control transformers with fused primary and secondary protection. • System overload trip, high temperature conditions or maintenance intervals for the pump will result in visual and audible alarms. |
|||||||||
Available Options |
N/A
• Oxygen assured pumps - prepared for use in WAGD systems. • Internal tank lining for corrosion resistance. • Variable Frequency Drive (VFD) Control Panel – with Premium NFPA controls. |
|||||||||
Note | N/A 3 Year Limited Warranty | |||||||||
|