Unit of Measure

Financing

Approve Payments

N/A Finance with Approve Payments

Dimensions

Length (A)

N/A 48 in

Width (B)

N/A 75 in

Height (C)

N/A 75 in

Inlet Connection Size

N/A 2 in

Outlet Connection Size

N/A 2 in

Tankmount Vacuum Systems

Pump Power

N/A 8.6 hp
Air Flow at 19 inHg1 N/A 56 scfm

Tank Capacity

N/A 240 gal

Tank Orientation

N/A Horizontal

Compressor Capacity

N/A 21,877 Btu/hr

Sound Level

N/A 79 dBA

System Full Load Current (F.L.A) at 208 V

N/A 62 A

System Full Load Current (F.L.A) at 230 V

N/A 60.8 A

System Full Load Current (F.L.A) at 460 V

N/A 32.6 A

System Weight

N/A 2,350 lb

Specifications

Electrical Panel Opening

N/A 42 in

Minimum Clearance for Ease of Maintenance

N/A 24 in

Motor Voltage

N/A 208 V230 V460 V

Phase

N/A 3

Power Supply Frequency

N/A 60 Hz

Vacuum System

N/A The Powerex medical vacuum system is designed to create a suction system to remove unwanted fluids or gases from hospital/ laboratory working areas. The medical vacuum system package is compliant with the NFPA 99 requirements for Risk Category 1 systems. Each system is completely tested before shipment and includes: • Multiple vacuum pumps and associated equipment. • AMSE air receiver. • Medical control panel. Each pump is factory piped to a common intake manifold. Vibration isolation pads are included with the system.

Rotary Vane Vacuum Pump

N/A Each pump shall be a dry rotary vane type vacuum pump, and shall be direct-driven by a C-face integrated electric motor enclosed in the pump housing and cooled by the pump fan. • Each vacuum pump shall be dry-running, featuring self-lubricating carbon/graphite vanes and shall not require any sealing fluid in the pumping chamber. • Each vacuum pump shall include an internal relief valve, a check valve, inlet and discharge flex connectors, a 5 micron inlet filter and a pump isolation valve.

Motor

N/A The motor is continuous duty, C-face, TEFC, suitable for 208-230, or 460V, 3 phase, 60 hertz electrical operation.

Receiver

N/A The system shall include an ASME rated air receiver. The tank shall be equipped with a vacuum gauge, a sight gauge, by-pass valves, and a manual drain. Optional internal tank lining for corrosion resistance.

Control Panel

N/A The control system provides automatic lead/lag sequencing and automatic alternation of all pumps in order to equalize the amount of usage among the available vacuum pumps. The Premium NFPA Control panel shall include a gateway server card and all features listed below: • PLC controller and a color touch screen panel which displays the operating status of the unit. • Building automation communication gateway with BacNet® protocol and Web server features. Web servers features include email notifications in case the system is in alarm or has achieved one its maintenance intervals and requires service. • Ethernet port for connection to BacNet® server or direct connection to facility Ethernet for viewing of system operations and status via device IP-address. • UL508A listed control panel in a NEMA 12 enclosure. The panel door will include: the HMI touch screen, an audible and visual alarms with an acknowledge button, and an HOA switch for each pump. • Magnetic starters. • Vacuum transducer for process control. • Single point power connection. • Redundant 120Vac control transformers with fused primary and secondary protection. • System overload trip, high temperature conditions or maintenance intervals for the pump will result in visual and audible alarms.

Note

N/A 3 Year Limited Warranty

  • 1 Air Flow shown at total system capacity.