The MATRIX medical gas verifier is designed for the real-time analysis and verification of hospital gases to ensure compliance with NFPA 99.
The MATRIX is a transportable instrument that is fully self-contained and able to operate on-site for 12 hours with no external support requirements. The real-time analysis data is displayed on the front LCD panels, which have integral alarm concentration indicators and loss of flow alarms.
The MATRIX Plus offers a data logging option which allows the analysis data to be displayed on a PC via a USB connection. This permits the real-time data to be displayed in a graphical plot. The data log can be archived with a location file name for use in reports. Additionally, the MATRIX Plus can be configured with a sampling pump for those applications which are not pressurized.
Product Highlights
Range (CO) |
N/A See Spec Sheet ppm |
Accuracy |
N/A See Spec Sheet |
Audible Alarm |
N/A 95 dB at 24 in |
Inlet Size |
N/A 1/4 ID Quick Release |
Operating Temperature Range |
N/A 32°F to 77°F ºF |
Power Requirements |
N/A Rechargeable Lithium Battery |
Weight |
N/A 19 lbs |
General Description |
N/A
Display: Alpha numeric LCD with Backlight Alarm: LED visual indicator per channel and flow alarm Operating Temperature: 32º to 77ºF (0º to 25ºC) recommended Humidity Range: 0-95% (non-condensing) Power: Rechargeable lithium battery & AC power cord Run Time: Up to 12 hours on a fully charged battery, or continuous with AC power cord Inlet Pressure for Dew Point: 55 psig (user provided regulator) Inlet Flow Rate for Sensors:0.5 liters/minute Enclosure: Polypropylene-copolymer case Size: 19 x 15 x 8 inches (48 x 38 x 20 cm) Weight: 19 lb. (8.6 kg) |