Compressed Air Systems, Inc.
9303 Stannum St
Tampa, FL, 33619-2658, US
Phone: (800) 626-8177
Fax: 813-628-0187
Email: [email protected]
Website: http://www.CompressedAirSystems.com
|

|
|


Item # ISA-60M, Enmet Ambient Air Monitor
|
Web Price QUOTE
|
|
|
The ISA-60M controller with remote MRI-5175 oxygen sensor/transmitter is specifically designed to continuously monitor for oxygen deficiency in a Magnetic Resonance Imaging (MRI) facility. Maintaining the quality of breathing air in an MRI room is an important safety consideration. Cryogenic gas used to maintain the magnetic field is highly dangerous if released into an enclosed space. Inert gas can create a hazardous situation by reducing the concentration of oxygen in the breathing air to unsafe levels. The ISA-60M system ensures the safety of the technicians and patients by continuously monitoring and displaying the levels of oxygen in the room and providing audio and visual alarms when there is an unsafe, low level oxygen condition in the area being monitored. Three versions available; Single Channel, Dual Channel, and Three Channel.
|
|
|
|
|
|
|
|
|
|
|
|
Temperature: 32° to 104°F (0° to 40°C) Relative Humidity: 10-99% RH, non-condensing Pressure, Atmospheric: 20 to 36 inHg (68 to 122 kPa) Approval: Safety Approvals: CE, Conforms to STD 61010-1, Certified to C22.2 #61010-1
ISA-60M MONITOR SPECIFICATIONS:
Standard Alarm Points: 19.5% & 17% decreasing oxygen Display: High-contrast, backlit Alarms: Visual: LEDs; Audible: Piezoelectric horn, 95 dB at 2 feet Alarm Relays: 5 programmable gas relays plus fault. All relays are programmable latching or non-latching, dry SPDT, 10 amps (resistive load only) at 110 VAC. Operating Power: 100 to 240 VAC and/or 12 VDC, 15 Watts Connections: Strain relief (2 supplied) for 0.20” to 0.35” OD Cable Operating Power: 100 to 240 VAC and/or 12 VDC, 15 Watts Enclosure: Thermoplastic box with clear, hinged front cover, designed for NEMA 12 and 4X Size: 10.5 H x 8.5 W x 5.8 D inches (27 x 22 x 15 cm) Weight: 8 lbs. (3.6 kg)
MRI-5175 TRANSMITTER:
Alarms: Visual: 3 Red LEDs; Visual Power/Fault: Green/Red LED; Optional Audible Alarm: 105 dB at 12 inches (30 cm) Outputs: 4-20 mA, RS-485 Modbus Power: 24 VDC from ISA-60M Connection to ISA-60M: Strain relief (1 supplied) for 24 VDC, 4-20 mA, 0.20” to 0.35” OD Cable Enclosure: Fiberglass-reinforced polyester with non-magnetic coating and hardware Dimensions: 5 x 5 x 4 inches (2 x 2 x 1.6 cm) Weight: 2 lbs. (0.9 kg) Oxygen Sensor supplied with MRI-5175: Type: Electrochemical; Range: 0 to 30.0% by Vol. O2; Resolution: 0.1% by Volume Oxygen; Typical Life: 4 to 5 years Sensor Cable from MRI-5175 to ISA-60M: 3-wire, shielded cable, typically 18 to 20 AWG, user supplied, UL Listed type CM, PLTC or TC
|
|
|
|
Product Highlights- Continuous monitoring of area with Long-life oxygen sensor
- Oxygen sensor/transmitter is designed for EMC compatibility
- User-programmable alarms and high capacity relays
- 4-20 mA inputs and outputs
- 100-240 VAC, 12 VDC power options
- Fault relay and visual fault alarm
- RS-232/485 with Modbus protocol
- Safety Approvals: CE
- Conforms to STD 61010-1
- Certified to C22.2 #61010-1
ENMET has a complete line of compressed airline monitoring and filtration systems for supplied breathing, respiratory air, process air, and hospital compressed airlines. There continuous in-line carbon monoxide CO monitors and three-stage respiratory airline filtration panels are designed to meet OSHA 1910.134 compressed breathing air monitoring requirements.
|
|
|
|
|
|
|
|
 |
|
|
|
|
|