Item # ISA-60M, Enmet Ambient Air Monitor

Web Price QUOTE

The ISA-60M controller with remote MRI-5175 oxygen sensor/transmitter is specifically designed to continuously monitor for oxygen deficiency in a Magnetic Resonance Imaging (MRI) facility. Maintaining the quality of breathing air in an MRI room is an important safety consideration. Cryogenic gas used to maintain the magnetic field is highly dangerous if released into an enclosed space. Inert gas can create a hazardous situation by reducing the concentration of oxygen in the breathing air to unsafe levels. The ISA-60M system ensures the safety of the technicians and patients by continuously monitoring and displaying the levels of oxygen in the room and providing audio and visual alarms when there is an unsafe, low level oxygen condition in the area being monitored. Three versions available; Single Channel, Dual Channel, and Three Channel.





Financing  · Specifications  · Power Requirements  · Dimensions  · General Description  · Additional Information

Financing

Approve Payments

Finance with Approve Payments



Specifications

Alarm Level Setting

Alarm 1: 5 ppm CO, user adjustable
Alarm 2: 10 ppm CO, user adjustable

Alarm 3: 20 ppm CO, user adjustable, All adjustable between 1 to 50 ppm

Audible Alarm

105 dB at 12 in
95 dB at 24 in

Inlet Size

Strain relief (2 supplied) for 0.20” to 0.35” OD Cable

Operating Temperature Range

32 to 104 ºF

Weight

10 lb


Power Requirements

Alternating Current (AC) Voltage

100 to 240 V

Direct Current (DC) Voltage

12 V

Power

15 W
12
W


Dimensions

Width

10.5 in
27
cm

Height

8.5 in
22
cm

Depth

5.8 in
15
cm


General Description
Temperature: 32° to 104°F (0° to 40°C)
Relative Humidity: 10-99% RH, non-condensing
Pressure, Atmospheric: 20 to 36 inHg (68 to 122 kPa)
Approval: Safety Approvals: CE, Conforms to STD 61010-1, Certified to C22.2 #61010-1

ISA-60M MONITOR SPECIFICATIONS:

Standard Alarm Points: 19.5% & 17% decreasing oxygen
Display: High-contrast, backlit
Alarms: Visual: LEDs; Audible: Piezoelectric horn, 95 dB at 2 feet
Alarm Relays: 5 programmable gas relays plus fault. All relays are programmable latching or non-latching, dry SPDT, 10 amps (resistive load only) at 110 VAC.
Operating Power: 100 to 240 VAC and/or 12 VDC, 15 Watts
Connections: Strain relief (2 supplied) for 0.20” to 0.35” OD Cable
Operating Power: 100 to 240 VAC and/or 12 VDC, 15 Watts
Enclosure: Thermoplastic box with clear, hinged front cover, designed for NEMA 12 and 4X
Size: 10.5 H x 8.5 W x 5.8 D inches (27 x 22 x 15 cm)
Weight: 8 lbs. (3.6 kg)

MRI-5175 TRANSMITTER:

Alarms: Visual: 3 Red LEDs; Visual Power/Fault: Green/Red LED; Optional Audible Alarm: 105 dB at 12 inches (30 cm)
Outputs: 4-20 mA, RS-485 Modbus
Power: 24 VDC from ISA-60M
Connection to ISA-60M: Strain relief (1 supplied) for 24 VDC, 4-20 mA, 0.20” to 0.35” OD Cable
Enclosure: Fiberglass-reinforced polyester with non-magnetic coating and hardware
Dimensions: 5 x 5 x 4 inches (2 x 2 x 1.6 cm)
Weight: 2 lbs. (0.9 kg)
Oxygen Sensor supplied with MRI-5175: Type: Electrochemical; Range: 0 to 30.0% by Vol. O2; Resolution: 0.1% by Volume Oxygen; Typical Life: 4 to 5 years
Sensor Cable from MRI-5175 to ISA-60M: 3-wire, shielded cable, typically 18 to 20 AWG, user supplied, UL Listed type CM, PLTC or TC



Additional Information
Product Highlights
  • Continuous monitoring of area with Long-life oxygen sensor
  • Oxygen sensor/transmitter is designed for EMC compatibility
  • User-programmable alarms and high capacity relays
  • 4-20 mA inputs and outputs
  • 100-240 VAC, 12 VDC power options
  • Fault relay and visual fault alarm
  • RS-232/485 with Modbus protocol
  • Safety Approvals: CE
  • Conforms to STD 61010-1
  • Certified to C22.2 #61010-1
ENMET has a complete line of compressed airline monitoring and filtration systems for supplied breathing, respiratory air, process air, and hospital compressed airlines. There continuous in-line carbon monoxide CO monitors and three-stage respiratory airline filtration panels are designed to meet OSHA 1910.134 compressed breathing air monitoring requirements.