Item # VVOTD0153, 1.5 HP Duplex Tankmount Oil-less Vane Open System VVOTD0153

Web Price QUOTE

Oil-Sealed Rotary Vane

Benefits:
  • Air-cooled design
  • Semi-synthetic oil
  • Oil-flooded for lubrication and heat dissipation
  • Lowest initial cost
  • Highest ultimate vacuum levels 29.3" Hg
  • No water and sewage costs
  • Most suitable for high ambient temperatures






Financing  · Dimensions  · Tankmount Vacuum Systems  · Specifications  · Vacuum System  · Rotary Vane Vacuum Pump  · Motor  · Receiver  · Control Panel  · Note  · Additional Information

Financing

Approve Payments

Finance with Approve Payments



Dimensions

Length (A)

53 in

Width (B)

34 in

Height (C)

87 in

Inlet Connection Size

1 in

Outlet Connection Size

1 in


Tankmount Vacuum Systems

Pump Power

1.5 hp

Air Flow at 19 inHg1

5.3 scfm

Tank Capacity

80 gal

Tank Orientation

Vertical

Compressor Capacity

3,819 Btu/hr

Sound Level

67 dBA

System Full Load Current (F.L.A) at 208 Volts

10.4 A

System Full Load Current (F.L.A) at 230 Volts

11 A

System Full Load Current (F.L.A) at 460 Volts

6.8 A

System Weight

550 lb
1 Air Flow shown at total system capacity.


Specifications

Brands

Powerex®

Vane Life

30,000 hours

Electrical Panel Opening

42 in

Minimum Clearance for Ease of Maintenance

24 in

Motor Voltage

208 V
230
V
460
V

Phase

3

Power Supply Frequency

60 Hz


Vacuum System
The Powerex medical vacuum system is designed to create a suction system to remove unwanted fluids or gases from hospital/ laboratory working areas. The medical vacuum system package is compliant with the NFPA 99 requirements for Risk Category 1 systems. Each system is completely tested before shipment and includes: • Multiple vacuum pumps and associated equipment. • AMSE air receiver. • Medical control panel. Each pump is factory piped to a common intake manifold. Vibration isolation pads are included with the system.



Rotary Vane Vacuum Pump
Each pump shall be a dry rotary vane type vacuum pump, and shall be direct-driven by a C-face integrated electric motor enclosed in the pump housing and cooled by the pump fan. • Each vacuum pump shall be dry-running, featuring self-lubricating carbon/graphite vanes and shall not require any sealing fluid in the pumping chamber. • Each vacuum pump shall include an internal relief valve, a check valve, inlet and discharge flex connectors, a 5 micron inlet filter and a pump isolation valve.



Motor
The motor is continuous duty, C-face, TEFC, suitable for 208-230, or 460V, 3 phase, 60 hertz electrical operation.



Receiver
The system shall include an ASME rated air receiver. The tank shall be equipped with a vacuum gauge, a sight gauge, by-pass valves, and a manual drain. Optional internal tank lining for corrosion resistance.



Control Panel
The control system provides automatic lead/lag sequencing and automatic alternation of all pumps in order to equalize the amount of usage among the available vacuum pumps. The Premium NFPA Control panel shall include a gateway server card and all features listed below:
  • PLC controller and a color touch screen panel which displays the operating status of the unit
  • Building automation communication gateway with BacNet® protocol and Web server features. Web servers features include email notifications in case the system is in alarm or has achieved one its maintenance intervals and requires service
  • Ethernet port for connection to BacNet® server or direct connection to facility Ethernet for viewing of system operations and status via device IP-address
  • UL508A listed control panel in a NEMA 12 enclosure. The panel door will include: the HMI touch screen, an audible and visual alarms with an acknowledge button, and an HOA switch for each pump
  • Magnetic starters
  • Vacuum transducer for process control
  • Single point power connection
  • Redundant 120Vac control transformers with fused primary and secondary protection
  • System overload trip, high temperature conditions or maintenance intervals for the pump will result in visual and audible alarms




Note
3 Year Limited Warranty



Additional Information
Benefit & Features
  • No oil anywhere
  • Long vane life
  • No water and sewage costs
  • Suitable for dedicated waste anesthesia systems
  • Expandable configurations for skid mounted systems up to quadplex
  • Fully compliant with NFPA 99