All Categories > Vacuum Pumps, Blowers & Parts > Medical, Laboratory and Industrial Vacuum Pump Systems > Medical Claw Enclosed Vacuum Systems > Item # MCED0507  


Item # MCED0507, 5 HP Duplex Medical Claw Enclosed System MCED0507

Web Price QUOTE

Powerex's trademarked sound enclosure systems offer the lowest noise levels in the industry. Each system is packaged in a rigid steel frame with powder coated panels and lined with sound-deadening insulation.





Financing  · Dimensions  · Tankmount Vacuum Systems  · Specifications  · Vacuum System  · Rotary Vane Vacuum Pump  · Motor  · Receiver  · Control Panel  · Available Options  · Additional Information

Financing

Approve Payments

Finance with Approve Payments



Dimensions

Length (A)

78 in

Width (B)

34 in

Height (C)

72 in

Inlet Connection Size

2 in

Outlet Connection Size

2 in


Tankmount Vacuum Systems

Pump Power

5 hp

Air Flow at 19 inHg1

33 scfm

Compressor Capacity

8,670 Btu/hr

Sound Level

64 dBA

System Full Load Current (F.L.A) at 208 Volts

33.4 A

System Full Load Current (F.L.A) at 230 Volts

31.8 A

System Full Load Current (F.L.A) at 460 Volts

16.4 A

System Weight

1717 lb
1 Air Flow shown at total system capacity.


Specifications

Brands

Powerex®

Electrical Panel Opening

42 in

Minimum Clearance for Ease of Maintenance

24 in

Motor Voltage

208 V
230
V
460
V

Phase

3

Power Supply Frequency

60 Hz


Vacuum System
The Powerex enclosed medical vacuum system is designed to create a suction to remove unwanted fluids or gases from hospital/laboratory working areas. The enclosed medical vacuum system package is compliant with the NFPA 99 requirements for Risk Category 1 systems.



Rotary Vane Vacuum Pump
Each pump shall be a rotary claw type vacuum pump, and shall be direct-driven through a shaft coupling by a C-face, TEFC electric motor.
  • Each vacuum pump shall be dry-running, featuring two claw type, non-contacting rotors and shall not require any sealing fluid in the pumping chamber, assuring virtually maintenance-free operation.
  • Each vacuum pump assembly shall include an integral relief valve.3
  • Each pump within the system shall include a check valve, inlet and discharge flex connectors, a 5 micron inlet filter and a pump isolation valve on the intake connection.




Motor
The motor is continuous duty, C-face, TEFC, suitable for 208-230, or 460V, 3 phase, 60 hertz electrical operation.



Receiver
The internally lined ASME rated vacuum receiver shall be equipped with a sight gauge, isolation valves, and a manual drain.



Control Panel
The control system provides automatic lead/lag sequencing and automatic alternation of all pumps in order to equalize the amount of usage among the available vacuum pumps. The Premium NFPA Control panel shall include a gateway server card and all features listed below:
  • PLC controller and a color touch screen panel which displays the operating status of the unit
  • Building automation communication gateway with BacNet® protocol and Web server features. Web server features include email notifications in case the system is in alarm or has achieved one its maintenance intervals and requires service.
  • Ethernet port for connection to BacNet® server or direct connection to facility Ethernet for viewing of system operations and status via device IP-address
  • UL508A listed control panel in a NEMA 12 enclosure. The panel door will include: the HMI touch screen, an audible and visual alarms with an acknowledge button, and an HOA switch for each pump
  • Magnetic contactors with Motor protector circuit breaks featuring through-the-door operators with lockable control knobs. Motor protector circuit breakers feature high inrush capability.
  • Vacuum transducer for process control
  • Single point power connection
  • Redundant 120Vac control transformers with fused primary and secondary protection
  • System overload trip, high temperature conditions or maintenance intervals for the pump will result in visual and audible alarms




Available Options
• Internal tank lining for corrosion resistance
• Variable Frequency Drive (VFD) Control Panel



Additional Information
The patented innovative sound enclosure reduces the vacuum pump sound intensity by 4-5 times which allows all Powerex WhisperVAC™ systems to safely fall in within the NIOSH 8-hour exposure limit. The reduced sound levels also allows for greater flexibility in system placement since there is less nuisance risk for patients and hospital staff.